Analytical HPLC System LAHS-B11

Analytical HPLC System LAHS-B11

Product Description

Analytical HPLC System LAHS-B11 is a unit equipped with LCD display UV-Visible wavelength range of 190 nm to 700 nm and double plunger reciprocating pump system for constant flow pump. It has gradient kit system for test result accuracy. Consists of separate workstation, sample injector and front mounted self-locking switch. Deuterium lamp is used as light source with valve mounting frame. It is ideally used in the detection of compounds in pharmaceutical industries, research institute, food industries, and petroleum industries, etc.catalog/Analytical-HPLC-System-LAHS-B11.pdf

    Share:

Available Range :

Pump Specifications : Solvent delivery : Double plunger series mode, floating plunger design, single stroke volume 30 µl Detector Specifications : Wavelength range : 190 to 700 nm Pump specification : Type : Quaternary Low Pressure Pump UV/VIS Detector : Wavelength Range : 190 to 800 nm Data-processing Workstation : System contain the tables for regular samples
Equipped with Internal and external standard calculation process
Automatic display, print, export of data and starting of another program on completion on measurement
Observe with the condition of 21 CFR part 11 directive of the FDA
Pump Specifications : Model No : HPP-B1 Detector Specifications : Model No : HPD-B2 High Pressure Constant Flow Pump : Double plunger reciprocating pump : 1 set UV/Visible Detector : Wavelength range : 190 to 700 nm Detector Specifications :
Model No HPD-A20 HPD-A21
Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm )
Wavelength accuracy ± 0.5 nm ± 2 nm
Wavelength repeatability 0.15 nm 0.3 nm
Band width 8 nm 8 nm
Optical path 10 nm 10 nm
Deuterium lamp 190 to 400 nm 190 to 400 nm
Tungsten lamp 400 to 800 nm 400 to 800 nm
Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5
Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4
Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8
Power 100 W 100 W
Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz
Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm
Weight 9.5 kg 9.5 kg
High Pressure Constant Flow Pump : Double plunger reciprocating pump : 2 sets UV/Visible Detector : Wavelength range : 190 to 700 nm
High Pressure Constant Flow Pump
Double plunger reciprocating pump2 sets
Flow range0.001 to 10.00 mL/min
Setting step size0.001 mL/min
Flow accuracy≤ ± 0.3 %
Flow stabilityRSD ≤ 00.1 %
Maximum pressure42 MPa
Display pressure error≤ 1.0 %
Pressure fluctuation≤ 1.0 %
Power consumption80 W
UV/Visible Detector
Wavelength range190 to 700 nm
Wavelength repeatability± 0.1 nm
Band width8 nm
Response time0 to 4.9 seconds
Linearity range1.8 AU (5 %)
Baseline Noise≤ ± 1 × 10-5AU
Baseline drift≤ 2.0 × 10-4AU/h
Minimum detectable concentration4.0 × 10-9 g/mL
Detector cellOptical path: 10 mm
Volume: 10 µL
Light sourceDeuterium lamp
Power consumption100 W
VoltageAC 110 V/ 220 V ± 10 %, 50/60 Hz
Packing Dimensions (W × D × H)600 × 440 × 380 mm
3 × (600 × 410 × 300 mm
Gross weight75 kg

  • Double plunger reciprocating pump system
  • Gradient system kits
  • Sample injector: 1 set
  • Deuterium lamp light source
  • Separate workstation
  • LCD display for value observation
  • Front mounted self-locking switch
  • Output: USB port
  • Compact and easy to install

Widely used in pharmaceutical, petroleum, research institute, food industrial applications in separation and detection of mixture of compounds in chemical analysis, poison analysis, environmental analysis, protein food detection, medicine testing, and polymer analysis.